RESEARCH LANDSCAPELAST VERIFIED 2026-08

Who measures gait, and with what.

What is actually used in sensor-based gait and physical-activity research. Every entry was verified against a primary source or the manufacturer's official page. Numbers carry their configuration — sampling rate, range, wearing site — because the same device at different settings is a different instrument. Scope is gait analysis only; cardiovascular measurement is out of scope here. Anything that could not be verified is simply not here.

01

What the field has reached, per task

This is not a leaderboard. Tasks that look alike are different problems once the reference, the cohort or the condition changes, so every figure carries its conditions at equal weight. Three things break comparison: the reference (INDIP's per-step truth, an examiner's stopwatch, a pedometer's daily total), the cohort (for frailty, prevalence moves the AUC on its own), and the condition (supervised versus free living, a single clip versus pooled).

Two studies hold the conditions fixed

The closest thing the field has to a benchmark: conditions fixed, method varied.

StudyHeld fixedResult
Kluge 2024 (JMIR Form Res 8:e50035)the same 83 participants across six disease groups, INDIP reference, ten gait-detection algorithmsthe wrist-versus-lower-back gap, quantified
Gao 2025 (J Gerontol A 80:glaf034)the same 1,282 participants, five step counters988 to 23,607 steps a day — a 24-fold spread

1.1Gait detection

ReportSiteCohortReferenceResult
Kluge 2024 — best on wrist (Brand 2022)wrist83, six disease groups, free livingINDIPsensitivity 0.55–0.81 / specificity 0.95–0.98 / gait-duration error 8.9–32.7%
Kluge 2024 — best on low back (Iluz 2014)low backibid.INDIPsensitivity 0.71–0.91 / specificity 0.96–0.99 / gait-duration error 6.3–23.5%
ElderNet (detection), Brand 2024wrist83 labelled—F1 0.829 ± 0.134
This workwrist225examiner-labelled walking boutsF1 0.922

The wrist loses to the lower back — 0.16 of sensitivity, on the same data. An arm moves independently of the legs, and that is the cost. Our own 0.922 is supervised structured walking, not the free living the three rows above are scored on.

1.2Step count

ReportCohortReferenceResult
Small 2024 (MSSE 56:1945)39 development / 30 externalannotated free-living datafree-living MAPE 12.5% (prior models 65–231%)
Watch Walk 2022101, 1,487 walksinstrumented walkwayMAPE 3.0–4.4% (supervised)
Gao 2025 (ARIC, five algorithms)the same 1,282method-to-method988 – 23,607 steps a day
Bai 2021 — Apple Watch 248, 24 hoursYamax pedometerMAPE 20.3%
Bai 2021 — Fitbit Charge 2ibid.ibid.MAPE 17.1%
This work (deployment)119, 4,179 clipsmanual per-step labelsMAE 1.684 steps per 10 s
This work (window given)111, 216 trialsibid.MAE 1.421 steps, contact F1 0.898

An order of magnitude separates supervised from free living: 3.0–4.4% on a walkway, 12.5% in the wild, and the latter is the deployment case. The consumer figures are daily totals; nobody has validated those devices step by step.

1.3Walking speed

MCID 10.0MDC 12.60481216ElderNet 20268.82Kirk 2024 (free living)11This work13.7
MAE (cm/s) — lower is betterper-step referencewalkway or lab instrumentstopwatch and coarserthis work
ReportSiteCohortReferenceResult
ElderNet 2026wristMobilise-D TVS, 18INDIPMAE 8.82 cm/s, ICC 0.87
Kirk 2024 (Mobilise-D)low back97INDIPlab 6–12 cm/s / free living 9–13 cm/s
This workwrist130examiner stopwatchMAE 13.7 cm/s

The reference matters: per-step truth above, a stopwatch below. The 4 m walk this would replace has a test-retest band of 10.8–14.4 cm/s, so 13.7 sits inside it.

1.4Step and stride length

MDC 7.004812Zadka 2024 (step length)6.08ElderNet 2026 (stride)11.61This work (stride)10.7
error (cm) — lower is betterper-step referencewalkway or lab instrumentstopwatch and coarserthis work
ReportSiteQuantityCohortResult
Zadka 2024 (npj Digit Med)low backstep length472, >80,000 stepsRMSE 6.08 cm (ICC 0.89) / 4.79 cm over ten steps
ElderNet 2026wriststride length18MAE 11.61 cm, ICC 0.72
This workwriststride length130MAE 10.7 cm, r 0.756

A stride is two steps, so 6.08 cm and 11.61 cm are not the same measurement — and one is worn at the waist, the other at the wrist.

1.5Cadence

MDC 8.40246810ElderNet 20263.11This work (100 clips pooled)2.13This work (single clip)8.08
MAE (steps/min) — lower is betterper-step referencewalkway or lab instrumentstopwatch and coarserthis work
ReportCohortResult
ElderNet 2026wrist, 18MAE 3.11 steps/min
This workwrist, 33, pooled 100 clips eachMAE 2.13 steps/min, r 0.951

Ours is the pooled figure. Per clip it is 8.08 and per subject 6.29 — a different condition from ElderNet's, and not directly comparable.

1.6Step-time variability

ReportMeasured withCohortResult
Hausdorff 2001pressure insoles52, followed one yearstride-time variability 106 ± 30 ms in fallers versus 49 ± 4 ms
This workwrist94r 0.633 against the reference, 65% of its own 0.975 reliability

No wrist-worn figure was found. Nothing in the searched literature estimates this quantity from the wrist and scores it against a reference; the only comparator is the clinical one.

1.7Frailty discrimination

demographics alone, this cohort — 0.7580.6000.7000.8000.900Cobo 20230.690This work (gait alone)0.726This work (with demographics)0.754Park 20210.769Song 20250.810Zheng 20250.920
AUC — higher is betterwristchest or thighunstatedthis work
ReportSiteCohortPrevalenceResult
Zheng 2025 (correction issued)—care-home residents—AUC 0.92
Song 2025 (conference abstract)thigh44—AUC 0.81
Park 2021chest pendant, 48 h25972%AUC 0.769 (0.60–0.74 by item)
UK Biobank 2025wrist10,156—mortality C 0.73 (Fried phenotype 0.72)
Cobo 2023wristNHANES, 1,279—AUC 0.69
This workwrist, free living21313.6%binary AUC 0.726 (gait alone) / 0.754 (with demographics), three-stage κw 0.309

0.769 at 72% prevalence and 0.726 at 13.6% are not the same measurement. And demographics alone reach 0.758 in this cohort — above Cobo's 0.69. Read any wrist-accelerometer figure by asking whether it was compared against age, sex and body size.

02

Research-grade raw recorders

Devices that store the sensor stream itself. Because nothing is reduced to step counts on-device, a different algorithm can be applied later — nearly every large cohort runs on these.

siteDeviceMakerSiteSensorsVerified configurationUsed by
AX3Axivity (UK)wrist3-axis accel100 Hz, ±8 g (UK Biobank setting)UK Biobank accelerometry sub-study (100k), CAPTURE-24
AX6Axivity (UK)wrist / low back6-axis (accel + gyro)—an option for Mobilise-D free-living recording
GT9X LinkActiGraph (US)wrist3-axis accel80 HzARIC (n = 1,282, mean age 83.4)
MoveMonitor+ (MM+)McRoberts (NL)low backIMU—Mobilise-D technical validation study (TVS)
Opal V2RAPDM / Clario (US)whole bodyaccel, gyro, magnetometer, barometer128 Hz for 12 h; up to 24 units synchronized per networkMobility Lab; maker reports use in 700+ peer-reviewed papers
activPALPAL Technologies (UK)thighaccel—Song 2025 (n = 44, 10 days); thigh placement separates sitting, standing and walking well

The wearing site decides what a device is good at: the low back for walking speed, the thigh for posture, the wrist for wear-time compliance. This project chose the wrist for the third reason.

03

Reference systems — what grades the wearables

The instruments the wearables are scored against — their precision is the ceiling of every evaluation. Whether a reference resolves individual steps in free living is what divides this field's difficulty.

siteSystemCompositionReference forSource
INDIP2 foot MIMUs, 1 low-back MIMU, 2 pressure insoles, 2 ankle distance sensorsthe only free-living reference in Mobilise-D; resolves individual initial contactsMobilise-D TVS
Marker-based motion captureopticalMobilise-D in-lab referenceibid.
GAITRiteinstrumented walkway (pressure floor)spatiotemporal parameters on a walkway; the source of the MDC figuresref. 7 (Youdas 2010, n = 24)
Yamax Digi-Walker SW-200mechanical pedometerthe baseline consumer step counts are compared againstref. 20 (Bai 2021, n = 48)

A supervised walkway is the easy case; per-step references in free living are as rare as Mobilise-D's INDIP. Consumer-device validation stops at daily totals against a pedometer.

04

Consumer wearables, as validated

For consumer devices the number that matters is the independent validation, not the spec sheet. Across 31 studies, step counting on the Apple Watch was validated in only 2 (ref. 21).

DeviceVerified accuracyConditionsSource
Apple Watch Series 2steps MAPE 20.3%n = 48, 24 h free living, pedometer referenceref. 20
Fitbit Charge 2steps MAPE 17.1%ibid.ref. 20
Fitbit Altasteps MAPE 20.7%ibid.ref. 20
Fitbit Charge / Charge HRMAPE <25% across 20 studiessystematic reviewref. 21 (Germini 2022)
05

Open analysis algorithms

Inseparable from the recorders: the same raw data under a different algorithm gives a different answer — on the same 1,282 subjects, daily step counts ranged 988 to 23,607, a factor of 24 (ref. 19).

NameWhat it isSource
ADEPT / Oak / SDT / Verisense / Stepcountwrist-acceleration step counting, all open sourceref. 19
ssl-wearables (OxWearables)weights pretrained on 700,000 person-days of UK Biobankref. 22
ElderNetself-supervised learning on free-living wrist data; separate gait-detection and gait-measure modelsrefs. 15, 16
06

Open datasets

The index — UK Biobank, CAPTURE-24, Mobilise-D TVS, ARIC, NHANES and more — is maintained as its own page.

Open the dataset index
07

Labs to watch

Groups that hold either the methods or the validation infrastructure of wearable gait measurement. Affiliations and roles were verified on official pages.

OxWearables / Wearables Group

Aiden Doherty — Oxford, Big Data Institute / Nuffield Dept. of Population Health

population-scale accelerometry and self-supervised learning; built UK Biobank's activity measures and CAPTURE-24

Center for the Study of Movement, Cognition and Mobility

Jeffrey M. Hausdorff — Tel Aviv Sourasky Medical Center / Tel Aviv University

gait variability, freezing of gait, falls; the origin of variability-based fall prediction, and a co-author of ElderNet

Brain and Movement Research Group (BAM)

Lynn Rochester (Mobilise-D coordinator) / Silvia Del Din — Newcastle University

regulator-grade validation of real-world gait; coordinates Mobilise-D across 34 institutions in 13 countries

Wearable & Implantable Technology (WIT)

Ciprian Crainiceanu / Vadim Zipunnikov — Johns Hopkins Bloomberg School of Public Health

statistical methods and open-source tooling for accelerometry; the ARIC / NHANES step-algorithm comparisons

Falls, Balance and Injury Research Centre / UNSW

Matthew Brodie / Kim Delbaere — NeuRA, Sydney

wrist-derived gait biomarkers (Watch Walk), validated on 78,822 UK Biobank participants, extended to fall and depression prediction

iCAMP

Bijan Najafi — Baylor College of Medicine

wearable frailty, falls and the diabetic foot; the prior art on scoring the five frailty items individually from sensors (ref. 10)

Politecnico di Torino, DET

Andrea Cereatti — Politecnico di Torino

INDIP, Mobilise-D's free-living reference system: foot MIMUs + pressure insoles + distance sensors

Center for Gerontology and Social Science / Frailty Research

島田裕之 / 荒井秀典 — NCGG (Japan)

the revised J-CHS criteria — the definition behind this project's training labels; also reports daily-gait monitoring versus frailty progression

They divide into method holders (Oxford, Johns Hopkins), validation-infrastructure holders (Newcastle / Mobilise-D, Politecnico di Torino) and clinical-question holders (Tel Aviv, Baylor, NCGG). Three lineages pursue gait from the wrist — Oxford, Tel Aviv and UNSW/NeuRA — and they are the ones this project must compare against directly.

REFERENCES
  1. 7Youdas JW, et al. Responsiveness of 2 procedures for measurement of temporal and spatial gait parameters in older adults. PM&R 2010;2(6):537–543.link
  2. 10Park C, Mishra R, Golledge J, Najafi B. Digital biomarkers of physical frailty and frailty phenotypes using sensor-based physical activity and machine learning. Sensors 2021;21(16):5289.link
  3. 15Continuous assessment of daily-living gait using self-supervised learning of wrist-worn accelerometer data. npj Digital Medicine (2026).link
  4. 16Brand YE, et al. Self-supervised learning of wrist-worn daily living accelerometer data improves the automated detection of gait in older adults. Scientific Reports 2024;14:20854.link
  5. 19Gao S, et al. Comparing step-counting algorithms for high-resolution wrist accelerometry data in older adults in the ARIC study. J Gerontol A 2025;80(6):glaf034.link
  6. 20Bai Y, et al. Comprehensive comparison of Apple Watch and Fitbit monitors in a free-living setting. PLoS One 2021;16(5):e0251975.link
  7. 21Germini F, et al. Accuracy and acceptability of wrist-wearable activity-tracking devices: systematic review. J Med Internet Res 2022;24(1):e30791.link
  8. 22Yuan H, et al. Self-supervised learning for human activity recognition using 700,000 person-days of wearable data. npj Digital Medicine 2024;7:91.link